Programmable DC Electronic Load Used in Fuel Cell Test System
HomeHome > Products > Electronic Load > Programmable DC Electronic Load Used in Fuel Cell Test System
Programmable DC Electronic Load Used in Fuel Cell Test System

Programmable DC Electronic Load Used in Fuel Cell Test System

Programmable DC Electronic Load Used in Fuel Cell Test SystemN62400 Series Low Voltage High Current DC Electronic Load P
Basic Info.
Model NO. N62400
Current Range 0-1200A
Operation Modes Cc,CV,Cp,Cr
Voltage Accuracy 0.05%+0.1%F.S.
Dynamic Mode up to 20kHz
Voltage Sense Four-Wire Remote Sense
Communication Interface LAN/RS232/Can
Transport Package Carton
Specification 3U, 19 inch
Trademark NGI
Origin China
HS Code 903084900
Production Capacity 50000/Year
Product Description
Programmable DC Electronic Load Used in Fuel Cell Test SystemN62400 Series Low Voltage High Current DC Electronic Load Product IntroductionN62400 series is developed based on NGI's years of experience in testing for fuel cell. It is with high accuracy, high reliability and high cost performance. N62400 series is with high power density and elegant appearance, which is available for benchtop use or installation in 19 inch rack.Application Fields1. Fuel cell test2. Other low voltage & high current applicationsMain Features*Min. operating voltage as low as 0.2V for full current 1200A*Stable and reliable CR/CP function supported by hardware*Loading high current at ultra-low voltage, convenient for fuel cell test*Built-in DCIR test function (Optional)*Dynamic mode up to 20kHz, editable rise and fall slew rate*Remote/local sense Power-off memory function*Programmable sequence testing function(SEQ), up to 100 groups sequence files, up to 50 steps per file*LCD screen for data display*Built-in auto test function, one-button operation for complex test tasks*Analog programming interface(APG), current monitoring interface, remote/local trigger function to realize complex function control and monitoring*Standard 19-inch, available for rack installation*Short-circuit simulationSettable Von/VoffThe Von latch function has two modes to meet your various test needs: enabled and disabled.

Programmable DC Electronic Load Used in Fuel Cell Test System

Programmable DC Electronic Load Used in Fuel Cell Test System


Exhibitions

Programmable DC Electronic Load Used in Fuel Cell Test System

Programmable DC Electronic Load Used in Fuel Cell Test System

Programmable DC Electronic Load Used in Fuel Cell Test System


Product Line

For more specification data sheets, please contact us.
ModelN62401-40-200N62402-40-400N62403-40-600
Voltage40V40V40V
Current200A400A600A
Power1000W2000W3000W
Min. Operating Voltage[email protected] [email protected][email protected] [email protected][email protected] [email protected]
CC Mode
Range0-20A0-200A0-40A0-400A0-60A0-600A
Setting Resolution1mA10mA1mA10mA1mA10mA
Setting Accuracy (23±5ºC)Low range: 0.1%+0.1%F.S., High range: 0.1%+0.15%F.S.
CV Mode
Range0-4V0-40V0-4V0-40V0-4V0-40V
Setting Resolution0.1mV1mV0.1mV1mV0.1mV1mV
Setting Accuracy (23±5ºC)0.05%+0.1%F.S.
CP Mode
Range0-1000W0-2000W0-3000W
Setting Resolution0.1W
Setting Accuracy (23±5ºC)0.5%+1%F.S.
CR Mode
Range0.004Ω-20Ω0.08Ω-200Ω0.002Ω-10Ω0.04Ω-100Ω0.002Ω-6.6Ω0.03Ω-66Ω
Setting Resolution16bits
Setting Accuracy (23±5ºC)0.35%+0.156S0.35%+15.6mS0.35%+0.312S0.35%+31.2mS0.35%+0.468S0.35%+46.8mS
Slew Rate
Current3.3-200A/ms200-1000A/ms6.6-400A/ms400-2000A/ms10-600A/ms600-3000A/ms
Voltage0.334-16.7V/ms16.7-166.7V/ms0.334-16.7V/ms16.7-166.7V/ms0.334-16.7V/ms16.7-166.7V/ms
Power3.3-200A/ms200-1000A/ms6.6-400A/ms400-2000A/ms10-600A/ms600-3000A/ms
Resistance3.3-200A/ms200-1000A/ms6.6-400A/ms400-2000A/ms10-600A/ms600-3000A/ms
Accuracy (23±5ºC)(1+35%)* Setting value
Voltage Measurement
Range0-4V0-40V0-4V0-40V0-4V0-40V
Readback Accuracy (23±5ºC)0.05%+0.1%F.S.
Current Measurement
Range0-20A0-200A0-40A0-400A0-60A0-600A
Readback Accuracy (23±5ºC)0.1%+0.1%F.S.
Power Measurement
Range0-1000W0-2000W0-3000W
Readback Accuracy (23±5ºC)0.5%+1%F.S.
Dynamic Mode
T1&T20.025-60ms1-60000ms0.025-60ms1-60000ms0.025-60ms1-60000ms
Resolution1μs1ms1μs1ms1μs1ms
Accuracy (23±5ºC)1μs+100ppm1ms+100ppm1μs+100ppm1ms+100ppm1μs+100ppm1ms+100ppm
Others
InterfaceLAN/RS232/CAN
AC InputSingle phase, 220V AC±10%, frequency 47Hz~63Hz
TemperatureOperating temperature: 0ºC~40ºC, storage temperature: -20ºC~60ºC
Operating TemperatureAltitude <2000m, relative humidity: 5%~90%RH(non-condensing), atmospheric pressure: 80~110kPa
Net WeightApprox. 20 kgApprox. 24 kgApprox. 28 kg
Dimension3U,132.5(H)*482.0(W)*612.0(D)mm